Tenders from Philippines Get free alerts
ITB No.063.26- 1 unit Portable Surface Ultrasound with Trolley
Country:
PhilippinesSector:
Medical equipments, pharmaceuticals and personal care productsDeadline:
26 Nov 2026AI Summary
The Philippine Heart Center is procuring a single, fully‑integrated portable surface ultrasound system equipped with a trolley for use in vascular access and superficial imaging. The unit must include a broadband high‑resolution linear array transducer capable of operating from 7.5 MHz to 18 MHz, with a single probe that can adjust between high‑frequency superficial and lower‑frequency deeper vessel imaging. The system shall feature a 11‑inch flat, medical‑grade monitor with 2 MP resolution (1024 × 728 pixels), light weight (2.4 kg) and a multi‑transducer module allowing simultaneous connectivity and quick switching among up to three probes, including small‑footprint hockey‑stick types. To support clinical workflow, the supply must provide all accessories and consumables—including thermal paper, gel, a black and white printer, 100 sterile probe covers, and gel tubes—as well as a backup unit for concurrent replacement during repairs.
Operational and environmental requirements mandate a three‑year warranty with bi‑annual maintenance and calibration during the warranty period, and the unit must register within ambient temperatures from 0 °C to 50 °C (humidity 15–90 %) for storage and operate comfortably between 20 °C and 30 °C with less than 70 % relative humidity. The power supply is specified as 100‑240 VAC, 50/60 Hz with a 19 V/3.4 A DC output, and the equipment must come integrated with Wi‑Fi, Bluetooth, Ethernet, USB ports, and cloud export capability, and be compatible with Android and iOS platforms for remote access. The product is required to hold ISO 13485 certification, comply with IEC 60601‑1 and related safety standards, have CE/UL/BIS or equivalent certification, and be FDA‑approved or provide evidence of a pending FDA application. Comprehensive user and maintenance training for PHC personnel, quarterly post‑market surveillance for one year, and documentation of calibration, inspection, and spare part lists must also be supplied.
All equipment must comply with IEC 60601‑1 standards.
Operational and environmental requirements mandate a three‑year warranty with bi‑annual maintenance and calibration during the warranty period, and the unit must register within ambient temperatures from 0 °C to 50 °C (humidity 15–90 %) for storage and operate comfortably between 20 °C and 30 °C with less than 70 % relative humidity. The power supply is specified as 100‑240 VAC, 50/60 Hz with a 19 V/3.4 A DC output, and the equipment must come integrated with Wi‑Fi, Bluetooth, Ethernet, USB ports, and cloud export capability, and be compatible with Android and iOS platforms for remote access. The product is required to hold ISO 13485 certification, comply with IEC 60601‑1 and related safety standards, have CE/UL/BIS or equivalent certification, and be FDA‑approved or provide evidence of a pending FDA application. Comprehensive user and maintenance training for PHC personnel, quarterly post‑market surveillance for one year, and documentation of calibration, inspection, and spare part lists must also be supplied.
All equipment must comply with IEC 60601‑1 standards.
Description
ITB No.063.26- 1 unit Portable Surface Ultrasound with Trolley
1 Description of Function
It is used to the surface imaging of superficial structures beyond the skin as, nerves, arteries, veins, adjacent muscles, ligament, and soft tissues. It safety aids in inserting invasive monitors.
2 Operational Requirements
2.1 System complete with all accessories as specified in the Technical Specifications below.
2.2 3 year warranty
2.3 Maintenance service every 6 months during warranty period and calibration
2.4 Availability of back up unit to replace defective unit while on-going repair
3. Technical Specifications
3.1 Ultrasound guidance for vascular access procedures should be performed during
broadband high-resolution linear array transducers
3.1.1 Frequencies between 7.5 and 18 MHz is preferred
3.1.2 Single probe that will cover this frequency adjustable for high frequency viewing
superficial vessel
3.1.3 This probe can also be adjustable to lower frequency for deeper vessel viewing
(look for lowest during pre-bid)
3.2 Small-footprint probes (including the “hockey stick” - type probe) may be helpful in
pediatric patients or in areas where space is limited, such as near the clavicle.
3.3 Color of spectral Doppler imaging may be helpful for identifying flow in vascular
structures and in aiding the differentiation of arteries and veins.
3.4 Above 10 inch. high resolution, flat pane; medical grade monitor 2 megapixel with wide
viewing angle and good color resolution.
3.4.1 Resolution: 1024 x 728 pixels are better.
3.4.2 Physical Dimensions: 11.6-inch high resolution screen, weight: 2.4 kg with the
kickstand, 319.6 x 223.2 x 31.8mm
3.4.3 Unit should be as per manufacturer's dimension with multi-transducer module
allowing simultaneous connectivity of 3 transducer switching from the software user
interface.
3.4.4 Wheel casters with breaks
4. System Configuration Accessories, spare and consumables
4.1 In addition to ultrasound equipment, certain additional equipment listed below
4.1.1 Thermal paper and ultrasound gel, black and white printer
4.1.2 100 pieces sterile probe covers
4.1.3 2 tubes of ultrasound gel
5. Environmental factors
4.1 The unit shall be capable of being stored continuously in ambient
temperature of 0-50 deg C and relative humidity of 15-90%
5.2 The unit shall be capable of operating in ambient temperature of
20-30 deg C and relative humidity of less than 70%
6. Power Supply
6.1 Power adapter AC input: 100-240 VAC, 50/60Hz fitted with PHC electrical outlet, 1.7 A (max). Adapter DC output: 19 V, 3.4 A
7. Standards, Safety and Training
7.1 Manufacturers/Supplier should have a valid ISO 13485 certificate of
Quality Management System for Medical devices.
7.2 The equipment shall be certified as compliant with IEC 60601-1, EN 60601-1,
ANSI/AAMI ES60601-1, Or other equivalent internationally recognized electrical safety
standards applicable to medical electrical equipment.
7.3 The equipment must conform to recognized and verifiable international or regional
standards, as evidenced by certification such as CE, UL or BIS or other equivalent
standards issued by accredited certification or regulatory bodies.
7.4 The product shall be FDA-approved/registered. If FDA registration has not yet been
issued, the bidder may submit proof of a pending FDA application, such as CMDN,
CMDR, or other equivalent FDA-Issued or FDA-acknowledgement document.
7.5 Comprehensive Technical Training for PHC BIOMED and User's training to staff for
familiarity with the system
7.6 The supplier must submit a certificate that it shall conduct PMS on a quarterly basis
for a minimum of one (1) year.
7.7 Shall meet IEC-60601-1-2: 2014 (Or Equivalent BIS) General Requirements of
Safety for Electromagnetic Compatibility or should comply with 89/336/EEC;
EMC-directive
8. Connectivity
8.1 Integrated Wi-fi, Bluetooth, and Ethernet for data transfer.
8.2 Compatibility with Android/iOS devices for remote access and monitoring
8.3 USB ports and cloud storage export capability
9. SLCC
9.1 Similar unit
10. Delivery
10.1 60 days
11. Distributorship
11.1 Autorized
12. Documentation
12.1 User/Technical/Maintenance manuals to be supplied in English.
12.2 Certificate of calibration and inspection.
12.3 List of important spare parts and accessories with their part number and costing.
12.4 List of Equipments available for providing calibration and routine maintenance
support as per manufacturer documentation in service/technical manual
Pre-bid Conference
Date : 22/10/2026
Time : 10:00 AM
Venue : Philippine Heart Center East Avenue, Quezon City 5th Floor, Old Chapel, Hospital Bldg., PHC
Reference Number: 13284090
Area of Delivery: Metro Manila
Trade Agreement: Implementing Rules and Regulations
Procurement Mode: Public Bidding
Classification: Goods
Category: Hospital / Medical Equipment
Bid Supplements: 0
Document Request List: 6
1 Description of Function
It is used to the surface imaging of superficial structures beyond the skin as, nerves, arteries, veins, adjacent muscles, ligament, and soft tissues. It safety aids in inserting invasive monitors.
2 Operational Requirements
2.1 System complete with all accessories as specified in the Technical Specifications below.
2.2 3 year warranty
2.3 Maintenance service every 6 months during warranty period and calibration
2.4 Availability of back up unit to replace defective unit while on-going repair
3. Technical Specifications
3.1 Ultrasound guidance for vascular access procedures should be performed during
broadband high-resolution linear array transducers
3.1.1 Frequencies between 7.5 and 18 MHz is preferred
3.1.2 Single probe that will cover this frequency adjustable for high frequency viewing
superficial vessel
3.1.3 This probe can also be adjustable to lower frequency for deeper vessel viewing
(look for lowest during pre-bid)
3.2 Small-footprint probes (including the “hockey stick” - type probe) may be helpful in
pediatric patients or in areas where space is limited, such as near the clavicle.
3.3 Color of spectral Doppler imaging may be helpful for identifying flow in vascular
structures and in aiding the differentiation of arteries and veins.
3.4 Above 10 inch. high resolution, flat pane; medical grade monitor 2 megapixel with wide
viewing angle and good color resolution.
3.4.1 Resolution: 1024 x 728 pixels are better.
3.4.2 Physical Dimensions: 11.6-inch high resolution screen, weight: 2.4 kg with the
kickstand, 319.6 x 223.2 x 31.8mm
3.4.3 Unit should be as per manufacturer's dimension with multi-transducer module
allowing simultaneous connectivity of 3 transducer switching from the software user
interface.
3.4.4 Wheel casters with breaks
4. System Configuration Accessories, spare and consumables
4.1 In addition to ultrasound equipment, certain additional equipment listed below
4.1.1 Thermal paper and ultrasound gel, black and white printer
4.1.2 100 pieces sterile probe covers
4.1.3 2 tubes of ultrasound gel
5. Environmental factors
4.1 The unit shall be capable of being stored continuously in ambient
temperature of 0-50 deg C and relative humidity of 15-90%
5.2 The unit shall be capable of operating in ambient temperature of
20-30 deg C and relative humidity of less than 70%
6. Power Supply
6.1 Power adapter AC input: 100-240 VAC, 50/60Hz fitted with PHC electrical outlet, 1.7 A (max). Adapter DC output: 19 V, 3.4 A
7. Standards, Safety and Training
7.1 Manufacturers/Supplier should have a valid ISO 13485 certificate of
Quality Management System for Medical devices.
7.2 The equipment shall be certified as compliant with IEC 60601-1, EN 60601-1,
ANSI/AAMI ES60601-1, Or other equivalent internationally recognized electrical safety
standards applicable to medical electrical equipment.
7.3 The equipment must conform to recognized and verifiable international or regional
standards, as evidenced by certification such as CE, UL or BIS or other equivalent
standards issued by accredited certification or regulatory bodies.
7.4 The product shall be FDA-approved/registered. If FDA registration has not yet been
issued, the bidder may submit proof of a pending FDA application, such as CMDN,
CMDR, or other equivalent FDA-Issued or FDA-acknowledgement document.
7.5 Comprehensive Technical Training for PHC BIOMED and User's training to staff for
familiarity with the system
7.6 The supplier must submit a certificate that it shall conduct PMS on a quarterly basis
for a minimum of one (1) year.
7.7 Shall meet IEC-60601-1-2: 2014 (Or Equivalent BIS) General Requirements of
Safety for Electromagnetic Compatibility or should comply with 89/336/EEC;
EMC-directive
8. Connectivity
8.1 Integrated Wi-fi, Bluetooth, and Ethernet for data transfer.
8.2 Compatibility with Android/iOS devices for remote access and monitoring
8.3 USB ports and cloud storage export capability
9. SLCC
9.1 Similar unit
10. Delivery
10.1 60 days
11. Distributorship
11.1 Autorized
12. Documentation
12.1 User/Technical/Maintenance manuals to be supplied in English.
12.2 Certificate of calibration and inspection.
12.3 List of important spare parts and accessories with their part number and costing.
12.4 List of Equipments available for providing calibration and routine maintenance
support as per manufacturer documentation in service/technical manual
Pre-bid Conference
Date : 22/10/2026
Time : 10:00 AM
Venue : Philippine Heart Center East Avenue, Quezon City 5th Floor, Old Chapel, Hospital Bldg., PHC
Reference Number: 13284090
Area of Delivery: Metro Manila
Trade Agreement: Implementing Rules and Regulations
Procurement Mode: Public Bidding
Classification: Goods
Category: Hospital / Medical Equipment
Bid Supplements: 0
Document Request List: 6
Get instant alerts for new Medical equipments, pharmaceuticals and personal care products tenders in Philippines.
Free email alerts — no account required.